HP-AFS/1 Automatic Immunoassay System whole blood POCT Vitamin D analyzer CE ISO
- Model
- HP083
Item specifics
- Type
- Immunoassay System
- Instrument classification
- Class II
- Function
- Disease Analysis
- Certificate
- CE ISO
Review
Description
HP-AFS/1 Automatic Immunoassay System whole blood POCT Vitamin D analyzer CE ISO
Quick Details
Model Number |
HP-AFS/1
|
||
Origin |
China
|
||
Instrument classification | Class II | ||
Application | Hosipital | ||
Function | Disease Analysis | ||
Certificate | CE ISO | ||
Language | English |
Product Description
Product Name
HP-AFS/1 Immunoassay system
HP-AFS Specific Protein Analyzer
is made up of light absorption test module, scattered light test module, fluorescence test module, press components, data transmission interface and printer. And the absorption test module is made up of absorption light path unit, and light-absorption sensor part. The scattered light test module is made up of scattered light path unit and scattered light sensor part. It can utilize nephelometry, fluorescence and absorbency methodologies to test the protein quantitatively in the samples. The model of the instrument (HP-AFS/1) is defined by the number of the test channel.
Principle
1, Nephelometric test principle
—Rayleigh theory of Scattering
When the particle diameter is more than the wavelength of the incident light, the light basically scatters forward.
—Rayleigh-Debye/Mie theory of Scattering
When the particle diameter is less than the wavelength of the incident light, the light scatters in all directions evenly
2, Fluorescence test principle
3, Absorbency test principle
Technical
Nephelometry
|
Linear range
|
100~4000 NTU
Linear correlation coefficient≥0.99 |
RSD
|
RSD≤8%
|
|
Stability
|
The turbidity variation≤10%
|
|
Accuracy
|
Relative deviation≤10%
|
|
Fluorescence | Linear range | 100~1600AU, Linear correlation coefficient r≥0.99 |
RSD | RSD≤8% | |
Stability | Fluorescence variation≤15% | |
Accuracy | Relative deviation≤10% | |
Sensitivity | Detection limit is not more than 2.5μmol/L | |
Absorbency | Linear range | See attached table 1 |
RSD | RSD≤1% | |
Stability | The absorbency deviation is not more than 0.005 at 340nm position within 20min | |
Table 1 | Linear range of absorbency | |
Linear range (A) | Deviation(%) | |
0.2~≤0.5 | +5-5 | |
>0.5~≤1.0 | +4-4 | |
>1.0~≤1.8 | +2-2 |